BDS Turnkey Dielectric / Impedance Spectrometers
without Temperature Control

Overview on BDS systems

  • BDS 50 and BDS 40
    These are our high end systems for the frequency range from 3 μHz - 40 (BDS 50) or 3 µHz to 20 MHz (BDS 40), covering 13.1 and 12.5 decades on the frequency scale, respectively. These systems are based on the Alpha-A modular measurement system for dielectric, conductivity, electrochemical and impedance spectroscopy with several test interfaces.

    The sample impedance is directly determined by the phase-sensitive measurement of the sample voltage and current at fixed frequency. 

    In combination with the ZGS active sample cell interface, the standard sample arrangement is a parallel plate capacitor with diameters from 40 mm to 10 mm and a spacing between the plates from 20
    mm to 10 μm. 

    For liquid samples, a vacuum sealed sample cell is offered, which prevents evaporation of sample material.

    Short Specification (applies at the sample)
    Frequency 3·10-5 .. 4.107 Hz
    Capacity 10-15
    .. 1 F
    Impedance 0.01 - 1014
    Ω
    Loss factor tan(δ) accuracy < 3·10-5, resolution 10-5

    System includes Alpha-ATB or Alpha-ANB mainframe with ZGS active sample cell interface, shielding unit, WinIMP software and data acquisition unit.

BDS 40

Detailed description including principles of operation, measurement and application examples (410 kb PDF)

 

Detailed specification charts (70 kb PDF)

BDS 40

  • BDS 70
    This equipment covers the high frequency range from 1 MHz to 3 GHz (8 GHz by option).  The sample impedance is determined by coaxial line reflectometry.

    The RF sample cell containing the parallel plate sample capacitor (diameter 12 mm - 3 mm, spacing 20
    mm - 3 mm) is mounted at the RF impedance analyzer APC-7 impedance port . The sample impedance determines the reflection factor of the sample capacitor, which is measured by the RF analyzer.

    Thin film samples down to 1
    μm can be measured in a special thin film cell. 

    If the dielectric sample cell is replaced by a magnetic sample cell, the complex permeability of materials can be measured, too. 

    Short Specification
    (applies at the sample)
    Frequency 106 .. 3*109 Hz
    Impedance 0.1 - 105 Ω
    Loss factor tan(δ) accuracy < 3*10-3

    System includes Agilent 4991 RF Impedance analyzer, BDS 2200 RF sample cell, WinIMP software and data acquisition unit.

     
    Note: BDS 70 can be used with other RF impedance or network analyzers at frequencies up to 8 GHz, too. Please contact us for details.

BDS 40

A more detailed description of BDS 60 with principles of operation and  measurement examples can be found in this Dielectrics Newsletter (450 kb PDF)   

An article about the thin film RF cell can be found in this Dielectrics Newsletter (320 kb PDF)   

BDS 40

  • Optional Alpha-A Test Interfaces for High Voltage, High Current, Electrochemical Impedance Spectroscopy and 4 Electrode Measurements

    These are options for systems based on an Alpha-A modular analyzer included in the BDS 40 and 80 systems.

Alpha-A+EC10