| POT/GAL
30V 2A Short Specification |
| Counter
Electrode Polarization |
Voltage
Dc accuracy
Resolution |
±30 Vp dc and / or ac
2 mV + 10-4 of value
1 mV dc, 20 μV ac |
Current
Dc accuracy
Dc resolution |
±2 A dc and / or ac
10-3 value + 3.10-4 range + 0.5pA
1/32768 current range, 0.1 pA min. |
| Power |
60 W continuous maximum |
| Output resistance |
1, 10, 100, 1k Ω |
| Bandwidth |
Dc .. 1 MHz |
Simultaneous voltage
and current limits |
Continuously adjustable
2 V .. 40 V and 2 mA .. 2.1 A |
| Current into shorted
load |
Continuous 2 A without
overheating |
|
| Reference
Voltage Inputs |
| Configurations |
Single or differential
configuration
with selectable driven shields |
Voltage Range
Dc accuracy
Dc resolution |
±30 Vp
200 μV + 10-4 of value
20 μV |
| Input impedance |
> 1012
Ω | 10 pF |
| Common mode
rejection |
< 10-4 below
100 kHz
< 10-3 below 1 MHz |
| Input bias current |
< 2 pA |
| Bandwidth |
Dc .. 10 MHz |
|
| Working
Electrode Current Input |
Current ranges
Dc accuracy
Dc resolution |
20 pA .. 2 A in factors of
10
10-3 value + 3.10-4 range + 0.5pA
10-5 of range, 0.05 pA min |
| Bandwidth |
Dc .. 10 MHz |
|
| Main
Control Loop |
| Operation modes |
Potentiostat, Galvanostat
and
Direct Voltage in combination with 2,
3 or 4 wire arrangement |
| Dc accuracy |
200 μV + 10-4
of value |
| Time constants |
300 ns .. 3 s in factors of
3.33 |
| Bandwidth |
Dc .. 10 MHz |
Electrolyte Rs
compensation |
Automatic Rs detection by
high frequency EIS, Rs compensation or correction |
|
| Impedance
Measurement |
| Frequency range |
3 μHz .. 1 MHz * |
| Impedance range |
10-4 .. 1013
Ω |
| Capacitance range |
1 fF ... 10 F |
|
Basic Accuracy |
|
Relative impedance,
Relative capacity,
Loss factor tan(δ)
|
< 3.10-4 ***
|
|
Phase Angle |
< 6 m° *** |
|
Resolution |
|
Relative impedance,
Relative capacity,
Loss factor tan(δ)
|
< 10-5 |
|
Phase angle |
< 0.6 m° |
|
| User Calibrations |
load, short, open, internal self
calibration and diagnostics |